ALANINE - ESR DOSIMETRY, FEASIBILITY AND POSSIBLE APPLICATIONS

Citation
F. Kuntz et al., ALANINE - ESR DOSIMETRY, FEASIBILITY AND POSSIBLE APPLICATIONS, Journal de chimie physique et de physico-chimie biologique, 95(4), 1998, pp. 796-799
Citations number
5
Categorie Soggetti
Biology,"Chemistry Physical
Volume
95
Issue
4
Year of publication
1998
Pages
796 - 799
Database
ISI
SICI code
Abstract
Alanine ESR dosimetry presents a great interest for quality controls i n radiotherapy. This new developped water Equivalent alanine dosimeter allows a reproducible dose measurement, by a non-destructif readout t echnique in a large dose range. In this paper the stability of the dos imeter response has been shown but also its independance with the ener gy or the dose rate of the absorbed radiation. Through this different studies, one can broaden the application field of alanine / ESR dosime try especially for in-vivo dosimetry. The results of the experiments a nd the intra operative treatment, indicate that this kind of dosimetry seems to be a promising technique for in-vivo quality controls in ele ctron beam, gamma ray or X ray radiotherapy.