A simple, closed-vessel microwave digestion method has been developed
for several polymer samples. A 0.15-g sample of high density polyethyl
ene, polystyrene, polyethylene/polypropylene blend, and polyethylene p
recursor was digested in 4 mL nitric acid in a high-pressure (75 bar)
closed-vessel digestion bomb. The digestate was filtered and analyzed
by inductively coupled plasma optical emission spectrometry (ICP-OES).
Depending on the polymer, 1-18 elements were determined in the three
sample types. Detection limits were in the low parts-per-billion range
, and precision was better than 5% relative standard deviation for mos
t metals. The accuracy of the method was determined by performing spik
e recoveries for 1-15 test elements. Recoveries were in the 90-100% ra
nge for all elements. The digestion technique eliminated the need for
harsh acid mixtures (including H2SO4 and HF) that are routinely used f
or polymer samples.