Rm. Lindsay et al., A COMPARISON OF METHODS FOR THE MEASUREMENT OF HEMODIALYSIS ACCESS RECIRCULATION - AN UPDATE, ASAIO journal, 44(3), 1998, pp. 191-193
The Grit-Line Model III (CRIT) was compared with the Transonics Hemodi
alysis Monitor (TRANS) with regard to its ability to measure access re
circulation (AR). Patients were studied during routine dialysis before
and after dialysis line reversal done to induce AR. A strong correlat
ion between GRIT and TRANS AR measurements was obtained (r = 0.95; p =
0.000; n = 58) and GRIT values were repeatable (normalized SD from me
an 7.82%). The CRIT values were not identical to TRANS; at low AR valu
es (< 20%) they overestimated, and at higher values they underestimate
d. Provided this is appreciated, GRIT AR values are clinically useful.