S. Mergemeier et al., BASIC EXPERIMENTAL STUDIES ON THE OPERATION OF PHOTOIONIZATION DETECTORS, Fresenius' journal of analytical chemistry, 361(1), 1998, pp. 29-33
Commercially available and self-made photoionization detectors have be
en studied in a simple flow manifold without chromatographic column. T
he self-made detector allowed the independent variation of the detecto
r volume, the electrode surface area and a selective irradiation of th
e anodes and cathodes. The photoionization of ethanol and acetone has
been measured in hydrogen as carrier gas. All experimental results wer
e consistent with a model based upon a photoionization proceeding in t
he adsorption layer of the electrodes as the main signal generating pr
ocess. It has been shown that the photoionization of molecules in the
gas phase plays only a marginal role.