S. Garbos et al., THE EFFECT OF PALLADIUM MODIFIER ON THE EFFICIENCY OF ANTIMONY HYDRIDE TRAPPING IN GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY (AAS), Fresenius' journal of analytical chemistry, 361(1), 1998, pp. 43-46
The performance of palladium as permanent chemical modifier electropla
ted on the surface of a graphite tube for the preconcentration of anti
mony hydride was examined and compared with thermally formed Pd-coatin
gs. The application of Pd-electroplated tubes allows to perform at lea
st 75 determination cycles without any significant change in the effic
iency of hydride collection. This is an advantage over the thermally f
ormed palladium coating which must be obtained individually before eac
h measurement. After the optimization of the system parameters a conce
ntration detection limit of 48 ng/L and an absolute detection limit of
71 pg for a 1.48 mi, sample were obtained. The procedure was applied
to the determination of antimony at a concentration level of 0.2 mu g/
L in a tap water sample.