SINGLE-USE MEDIATOR ELECTROCHEMICAL BIOSENSORS FOR GLUCOSE ASSAY

Citation
Va. Bogdanovskaya et al., SINGLE-USE MEDIATOR ELECTROCHEMICAL BIOSENSORS FOR GLUCOSE ASSAY, Russian journal of electrochemistry, 33(4), 1997, pp. 423-426
Citations number
9
Categorie Soggetti
Electrochemistry
ISSN journal
10231935
Volume
33
Issue
4
Year of publication
1997
Pages
423 - 426
Database
ISI
SICI code
1023-1935(1997)33:4<423:SMEBFG>2.0.ZU;2-W
Abstract
The problems related to the development of single-use biosensors inten ded for glucose assay are considered. A selective sensor is designed o n the basis of carbon materials, an enzyme (glucose oxidase), and a me diator (benzoquinone). The response current is shown to be a linear fu nction of the glucose concentration at concentrations ranging from 3 t o 30 mM. Reproducibility of the biosensor is equal to 8-10%.