S. Mesli et al., EVALUATION OF THE NEW AUTOMATED ELISA VIDAS(R) LP(A) ASSAY - COMPARISON WITH AN IMMUNONEPHELEMETRIC METHOD, Clinica chimica acta, 272(2), 1998, pp. 183-195
Lipoprotein (a) [Lp(a)] is a lipoprotein which has a plasma concentrat
ion that is highly correlated with cardiovascular disease. In this stu
dy, the new Lp(a) assay for the Vitek Immuno-Diagnostic Assay System (
VIDAS(R)) developed by bioMerieux was evaluated. This method uses an e
nzyme linked fluorescent immunoassay (ELFIA) technique. Within-run and
between-run reproducibility of the Vidas(R) Lp(a) assay are character
ized by coefficients of variation (CV) of 3 to 5.9% and 3.9 to 5.9%, r
espectively. Using analysis of variance, no statistical difference was
shown between ELFIA and immunonephelometric assay (INA). When compari
ng results of the Vidas(R) Lp(a) test with the INA, a highly significa
nt correlation of r = 0.9708 and regression line equation y = 0.963x -
0.037 was found. Interference of lipemia was studied: no influence wa
s observed up to 12.3 mmol l(-1) triglycerides. No interference of hae
moglobin was noted for Lp(a) > 0.20 g l(-1). Hyperbilirubinemia (> 120
mu mol l(-1)) led to results being underestimated for concentrations
of Lp(a) which were < 0.20 g l(-1). No pre-analytical interference of
citrate was measured but pre-analytical interference of EDTA was found
. In conclusion, this new fully automated immunofluorimetric Lp(a) ass
ay enables to the rapid, accurate and reliable determination of Lp(a)
in blood samples. (C) 1998 Published by Elsevier Science B.V. All righ
ts reserved.