CALIBRATION OF A FIBER TIP ULTRASONIC SENSOR UP TO 50 MHZ AND THE APPLICATION TO SHOCK-WAVE MEASUREMENT

Citation
C. Koch et al., CALIBRATION OF A FIBER TIP ULTRASONIC SENSOR UP TO 50 MHZ AND THE APPLICATION TO SHOCK-WAVE MEASUREMENT, Ultrasonics, 36(1-5), 1998, pp. 721-725
Citations number
18
Categorie Soggetti
Acoustics,"Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
0041624X
Volume
36
Issue
1-5
Year of publication
1998
Pages
721 - 725
Database
ISI
SICI code
0041-624X(1998)36:1-5<721:COAFTU>2.0.ZU;2-C
Abstract
An optical calibration technique is described, which allows the invest igation of the frequency response of ultrasonic detectors up to a freq uency of 50 MHz. This technique is applied to determine the transfer f unction of a metal-coated ultrasonic fiber tip sensor which is based o n an interferometric measurement of the displacement using a Michelson interferometer. The transfer function is used to correct shock wave d ata which are, by this means, measured with a calibrated system. (C) 1 998 Elsevier Science B.V.