H. Suzuki et al., PROBLEMS ASSOCIATED WITH THE THIN-FILM AG AGCL REFERENCE ELECTRODE AND A NOVEL STRUCTURE WITH IMPROVED DURABILITY/, Sensors and actuators. B, Chemical, 46(2), 1998, pp. 104-113
We have developed a novel thin-film Ag/AgCl reference electrode which
shows dramatic improvement in its durability. To make the electrode re
sistant to heat treatment in its fabrication, a metal adhesive layer i
n the currently used thin-film Ag/AgCl electrodes was replaced by a po
lyimide layer, and a gold backbone structure was employed. Unlike the
currently used structure, the entire surface of the silver layer was c
overed with a polyimide layer and AgCl was grown from the periphery of
the pattern. The structure was resistant to baking at 300 degrees C.
Whereas the currently used Ag/AgCl electrode has a very short lifespan
in concentrated KCl solutions, the novel electrode showed surprising
resistance in these solutions. In a KCl and AgCl saturated solution, t
he novel electrode gave a stable potential for approximately 8 h, whil
e the currently used electrode functioned for only 1.5 min. A dramatic
difference was also observed in the resistance to interfering materia
ls. Although 1 mM Kl or K2S was enough to break the currently used ele
ctrode, the novel electrode was resistant to these materials and showe
d much smaller potential shift. In the novel Ag/AgCl structure, the ra
te of AgCl formation was significantly delayed and the spread of the A
gCl layer was visualized. Therefore, any of the electrochemical and ch
emical methods of AgCl formation could be easily applied. (C) 1998 Els
evier Science S.A. All rights reserved.