CARDIAC MR FLOWMETRY - EXPERIMENTAL VALID ATION AND RESULTS IN PATIENTS WITH OPERATED CARDIAC DEFICIENCIES

Citation
N. Hosten et al., CARDIAC MR FLOWMETRY - EXPERIMENTAL VALID ATION AND RESULTS IN PATIENTS WITH OPERATED CARDIAC DEFICIENCIES, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 168(5), 1998, pp. 480-487
Citations number
15
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren
ISSN journal
09366652 → ACNP
Volume
168
Issue
5
Year of publication
1998
Pages
480 - 487
Database
ISI
SICI code
0936-6652(1998)168:5<480:CMF-EV>2.0.ZU;2-0
Abstract
Purpose: A flow-sensitive MR sequence (phase-contrast technique) was e valuated in phantom studies with regard to factors influencing measure ments and correctness of results. The sequence was additionally used f or functional evaluation of operated congenital heart disease. Methods : Pulsatile and constant flow were produced with the help of a phantom . Influence of angulation, range and vessel bending was evaluated. An examination protocol was developed from the results. 35 patients with surgically repaired congenital heart disease or without repair were ex amined. Results: A range preset below the actual flow velocity as well as angulation of more than 20 degrees were isolated as main pitfalls in MR flowmetry. In addition to morphological MR findings flow measure ments were possible in 11 patients at vessel sites which were not or n ot completely suited for examination by Doppler ultrasound. Conclusion : The evaluated phase-contrast technique allows for fast and reliable flow quantification if the influences identified in phantom studies ar e considered.