EVALUATION OF A MICROCONCENTRIC NEBULIZER-CYCLONIC SPRAY CHAMBER FOR FLOW-INJECTION SIMULTANEOUS MULTIELEMENT INDUCTIVELY-COUPLED PLASMA OPTICAL-EMISSION SPECTROMETRY

Citation
Td. Hettipathirana et De. Davey, EVALUATION OF A MICROCONCENTRIC NEBULIZER-CYCLONIC SPRAY CHAMBER FOR FLOW-INJECTION SIMULTANEOUS MULTIELEMENT INDUCTIVELY-COUPLED PLASMA OPTICAL-EMISSION SPECTROMETRY, Journal of analytical atomic spectrometry, 13(5), 1998, pp. 483-488
Citations number
41
Categorie Soggetti
Spectroscopy
ISSN journal
02679477
Volume
13
Issue
5
Year of publication
1998
Pages
483 - 488
Database
ISI
SICI code
0267-9477(1998)13:5<483:EOAMNS>2.0.ZU;2-8
Abstract
Merits of a micro concentric nebuliser and a cyclonic spray chamber fo r sample introduction to simultaneous multielement inductively coupled plasma optical emission spectrometry (ICP-OES) were evaluated. Sample delivery was both continuous and by flow injection (FI), All signal p rofiles were acquired and processed in the time-resolved mode. At solv ent how rates of 0.1 and 1.0 ml min(-1), solvent transport efficiency to the ICP is about 90 and 50%, respectively, As a consequence, improv ements in detection limits were observed compared with those using a c onventional concentric nebuliser. Application of Sc as an internal sta ndard to improve precision with FI sample delivery was also studied.