HEMOLYSIS ASSOCIATED WITH CONTINUOUS VENOVENOUS RENAL REPLACEMENT CIRCUITS

Citation
P. Bierer et al., HEMOLYSIS ASSOCIATED WITH CONTINUOUS VENOVENOUS RENAL REPLACEMENT CIRCUITS, Anaesthesia and intensive care, 26(3), 1998, pp. 272-275
Citations number
13
Categorie Soggetti
Anesthesiology,"Emergency Medicine & Critical Care
ISSN journal
0310057X
Volume
26
Issue
3
Year of publication
1998
Pages
272 - 275
Database
ISI
SICI code
0310-057X(1998)26:3<272:HAWCVR>2.0.ZU;2-1
Abstract
Extracorporeal circuits can cause haemolysis resulting in an increase in plasma-free haemoglobin (PFHb). High pressures and clots within the circuit have been identified as factors increasing the likelihood of haemolysis. continuous venovenous haemodiafiltration (CVVHD) is associ ated with high circuit pressures as the pump-driven circuit clots over a period of time. PFHb was measured during CVVHD to determine of circ uit life,maximum circuit pressure or the clotting of the haemofilter w as associated with evidence of haemolysis. Circuit life up to 50 hours , circuit pressures or haemofilter clotting had not significant effect on PFHb. There was a small rise in PFHb in the circuits lasting beyon d 50 hours. CVVHD circuits can be run up to 50 hours without concern f or haemolysis.