Abstract
A homogeneous fluorophore-linked assay was used to develop a postcolum n reaction detection system for high-performance liquid chromatography (HPLC). Biotin and biocytin were chosen as the model analytes. The ef fluent from the HPLC column was merged with a reagent stream containin g avidin that was labeled with fluorescein isothiocyanate (avidin-FITC ). The binding of the separated analytes by the labeled avidin was acc ompanied by an enhancement of the fluorescence intensity at 520 nm. Th is increase in fluorescence was proportional to the concentration of t he analytes and constituted the analytical signal. The procedure was o ptimized with respect to the reagent concentration and the flow-rate o f the reagent solution. Analytical characteristics of the method were determined. The procedure was highly selective for biotin and its deri vatives. The detection limits for biotin and biocytin were 89 and 94 p g, respectively, for 20-mu l injections. The developed postcolumn reac tion detection system was validated by determining biotin in a liquid vitamin preparation and a horse-feed supplement.