VALIDATED METHOD FOR THE DETERMINATION OF IDAZOXAN IN HUMAN PLASMA BYLIQUID-CHROMATOGRAPHY WITH TANDEM MASS-SPECTROMETRIC DETECTION

Citation
Nc. Vandemerbel et al., VALIDATED METHOD FOR THE DETERMINATION OF IDAZOXAN IN HUMAN PLASMA BYLIQUID-CHROMATOGRAPHY WITH TANDEM MASS-SPECTROMETRIC DETECTION, Journal of chromatography B. Biomedical sciences and applications, 708(1-2), 1998, pp. 113-120
Citations number
6
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical sciences and applications
ISSN journal
13872273 → ACNP
Volume
708
Issue
1-2
Year of publication
1998
Pages
113 - 120
Database
ISI
SICI code
0378-4347(1998)708:1-2<113:VMFTDO>2.0.ZU;2-E
Abstract
A liquid chromatography-tandem mass spectrometry method for the determ ination of idazoxan in human (heparin) plasma is presented, which was developed and validated using 500 mu l of sample. Sample preparation c onsisted of the addition of fluoroidazoxan as the internal standard, e xtraction at alkaline conditions into tert.-butyl methyl ether, follow ed by centrifugation, evaporation of the solvent and reconstitution in methanol. After a short chromatographic run, detection took place by ionspray tandem mass spectrometry in positive ion mode. Validation res ults on linearity, specificity, accuracy, precision and stability, as well as application of the method to samples from a clinical trial, ar e shown. The validated calibration range is from 0.300 to 100 ng/ml, w ith accuracy (bias) and precision (coefficient of variation) being bel ow 15% at all levels. A sample throughput of, typically, 150 per day c an be achieved. (C) 1998 Elsevier Science B.V.