H. Jalink et al., MEASUREMENT OF RADIATION-INDUCED LOCALIZED ABSORPTION IN CHROMATOGRAPHIC ELUENTS BY MEANS OF THE CO LASER AND THE PZT PYROELECTRIC DETECTOR, Instrumentation science & technology, 26(2-3), 1998, pp. 305-313
Combined use of the CO laser and PZT piezoelectric detector is propose
d as a sensitive and specific detector for gas chromatography. The fea
sibility of this detection scheme is demonstrated for a model mixture
containing varying proportions of acetone in water. After separation o
n a GC column, spectra of both constituents were recorded in the 1600
cm(-1) to 1900 cm(-1) range that encompasses the characteristic freque
ncy of C=O bond in acetone. Reasonably high sensitivity (mass concentr
ation limit is 1.2x10(-8) g acetone) and a good degree (factor 530) of
spectral discrimination was achieved by a proper selection of analyti
cal wavelengths.