J. Ayrton et al., USE OF GENERIC FAST GRADIENT LIQUID-CHROMATOGRAPHY TANDEM MASS-SPECTROSCOPY IN QUANTITATIVE BIOANALYSIS, Journal of chromatography B. Biomedical sciences and applications, 709(2), 1998, pp. 243-254
Citations number
13
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical sciences and applications
Short narrow analytical HPLC columns have been used successfully with
high linear flow-rates and combined with mass spectrometric detection
to produce a generic approach to quantitative bioanalysis. The approac
h has been used to validate several assays in the low ng/ml region and
an example is given in this paper. When combined with a simple solid-
phase extraction process the need for complicated, time consuming meth
od development has been removed for the majority of pharmaceutical com
pounds. The approach takes advantage of not only the extra selectivity
of the MS-MS detector but the excellent resolution and peak shape pro
duced by gradient elution. (C) 1998 Elsevier Science B.V. All rights r
eserved.