Ji. Brodfuehrer et al., ACHIRAL AND CHIRAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHODS FOR CLINAFLOXACIN, A FLUOROQUINOLONE ANTIBACTERIAL, IN HUMAN PLASMA, Journal of chromatography B. Biomedical sciences and applications, 709(2), 1998, pp. 265-272
Citations number
5
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical sciences and applications
Achiral and chiral HPLC methods were developed for clinafloxacin, a qu
inolone antimicrobial agent. For achiral assay, analytes were isolated
from plasma by precipitating plasma proteins. Separation was achieved
on a C-18 column using an isocratic eluent of ion pairing solution-ac
etonitrile (80:20, v/v) at 1.0 ml/min with UV detection at 340 nm. The
ion pairing solution was 0.05 M citric acid, 1.15 mM tetrabutylammoni
um hydroxide and 0.1% ammonium perchlorate. Inter-assay accuracy was w
ithin 4.9% with an inter-assay precision of 3.7% over a quantitation r
ange of 0.025 to 10.0 mu g/ml. For chiral assay, analytes were isolate
d from plasma by solid-phase extraction. Separation was achieved on a
Crownpak CR(I) column using an isocratic eluent of water-methanol (88:
12, v/v) containing 0.1 mM decylamine at 1.0 ml/min with UV detection
at 340 nm. Perchloric acid was added to adjust pH to 2. Inter-assay ac
curacy was within 3.5% with a inter-assay precision of 5.4% over a qua
ntitation range of 0.040 to 2.5 mu g/ml. (C) 1998 Elsevier Science B.V
. All rights reserved.