V. Vinhhung et al., USE OF A SIMULATOR WITH CT OPTION IN RADIOTHERAPY OF MACULAR DEGENERATION, International journal of radiation oncology, biology, physics, 41(3), 1998, pp. 721-727
Citations number
18
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Purpose: To assess the accuracy of a conventional simulation procedure
in radiotherapy of age-related macular degeneration, Methods and Mate
rials: A computed tomographic (CT) extension attached to the treatment
simulator was used to acquire CT images immediately after conventiona
l simulation in 18 patients referred for treatment of age-related macu
lar degeneration. Analysis was performed on 16 one-sided treatment cas
es for whom images were obtained. Error was estimated by the displacem
ent between the observed treatment isocenter and the intended isocente
r based on reconstructed eye geometry. Results: Based on single slice
measurements, the mean error amplitude was 23 mm (range 0.2-5.6). Base
d on three dimensional eye globe reconstruction, the mean error amplit
ude was 2.8 mm (range 0.8-5.3). An incidental finding previously unrep
orted was the lower image quality at the center of the simulator-CT im
age acquisition held. Conclusions: Small but significant errors from c
onventional simulation were noted. The integrated simulation-CT proced
ure may help correct the errors to improve the accuracy of simulation
setup. The lower image quality at the tenter of image acquisition fiel
d requires adaptation of the simulation-CT procedure. (C) 1998 Elsevie
r Science Inc.