DETERMINATION OF ALUMINUM AND ITS FLUORO COMPLEXES IN NATURAL-WATERS BY ION CHROMATOGRAPHY

Citation
S. Motellier et H. Pitsch, DETERMINATION OF ALUMINUM AND ITS FLUORO COMPLEXES IN NATURAL-WATERS BY ION CHROMATOGRAPHY, Journal of chromatography, 660(1-2), 1994, pp. 211-217
Citations number
21
Categorie Soggetti
Chemistry Analytical
Journal title
Volume
660
Issue
1-2
Year of publication
1994
Pages
211 - 217
Database
ISI
SICI code
Abstract
Ion chromatography was applied to the determination of aluminium and i ts fluoro complexes in natural waters. The separation was carried out on a cation-exchange column. The aluminium species were detected by po stcolumn reaction with Tiron followed by UV spectrophotometry. The met hod requires the adjustment of the pH and ionic strength of the sample to those of the mobile phase immediately prior to injection. Al3+, Al F2+ and AlF2+: are eluted separately while all hydroxo complexes are r eadily dissociated and eluted along with Al3+ under these conditions. The sum of peak areas, which represents the total aluminium concentrat ion, was conserved whatever the amount of fluoride in the sample. Line arity of calibration was observed over the range 20-2000 mu g l(-1). F urther, the speciation of fluoro-aluminium complexes as determined exp erimentally by ion chromatography is in good agreement with calculatio ns based on complexation constants. The applications and limitations o f the method are discussed.