D. Cavouras et al., AN INVESTIGATION OF THE IMAGING CHARACTERISTICS OF THE Y2O2S-EU3-RAY-DETECTORS OF DIGITAL MAMMOGRAPHY( PHOSPHOR FOR APPLICATION IN X), Applied radiation and isotopes, 49(8), 1998, pp. 931-937
Citations number
22
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging","Chemistry Inorganic & Nuclear
Y2O2S:Eu laboratory prepared screens were evaluated as mammographic im
age receptors and were compared to similarly prepared screens of Gd2O2
S:Tb and Y2O2S:Tb phosphor materials, often used in X-ray imaging dete
ctors. The evaluation was performed by determining the Modulation Tran
sfer Function (MTF) and the spatial frequency dependent Detective Quan
tum Efficiency (DQE). Y2O2S:Eu exhibited higher DQE values at low freq
uencies and given its good spectral matching with digital optical dete
ctors, it may be appropriate for use in X-ray digital mammography. (C)
1998 Elsevier Science Ltd. All rights reserved.