Fc. Camargo et al., A FLOW-INJECTION SYSTEM FOR SPECTROPHOTOMETRIC CATALYTIC DETERMINATION OF COPPER IN PLANT DIGESTS AND NATURAL-WATERS, Analytical sciences, 14(3), 1998, pp. 565-569
The Cu(II)-catalyzed oxidation of 2,6-dichlorophenolindophenol (DCPI)
in the reduced form by hydrogen peroxide was exploited for the develop
ment of a flow injection procedure for copper determination in plant d
igests and natural waters. For selectivity enhancement, a Chelex-100 r
esin mini-column (50-100 mesh) was incorporated to the manifold. The p
roposed system is very stable, handles about 60 samples per hour, yiel
ds precise results (RSD<2% within 5.0 and 70.0 mu g l(-1) Cu) and requ
ires only 0.87 mg DCPI per measurement. Beer's law is followed up to a
bout 100 mu g l(-1) Cu and the detection limit is 0.09 mu g l(-1) Cu (
in the injectate). Baseline drift was not observed during extended (5
h) operation periods. Results were in agreement with certified values
of standard reference materials and recoveries within 95 and 104% were
found.