HIGH-SPATIAL-RESOLUTION GAMMA-IMAGING DETECTOR BASED ON A 5''-DIAMETER R3292 HAMAMATSU PSPMT

Citation
R. Wojcik et al., HIGH-SPATIAL-RESOLUTION GAMMA-IMAGING DETECTOR BASED ON A 5''-DIAMETER R3292 HAMAMATSU PSPMT, IEEE transactions on nuclear science, 45(3), 1998, pp. 487-491
Citations number
7
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
45
Issue
3
Year of publication
1998
Pages
487 - 491
Database
ISI
SICI code
0018-9499(1998)45:3<487:HGDBOA>2.0.ZU;2-P
Abstract
High resolution imaging gamma-ray detectors were developed using Hamam atsu's 5 '' diameter R3292 position sensitive PMT (PSPMT) [1] and a va riety of crystal scintillator arrays. Special readout techniques were used to maximize the active imaging area while reducing the number of readout channels. Spatial resolutions approaching 1 mm were obtained i n a broad energy range from 20 to 511 keV. Results are also presented of coupling the scintillator arrays to the PMT via imaging light guide s consisting of acrylic optical fibers.