A METHOD FOR COMPARING DEGRADATION OF BORON-TRIFLUORIDE AND HELIUM DETECTORS IN NEUTRON AND GAMMA-FIELDS

Citation
T. Qian et al., A METHOD FOR COMPARING DEGRADATION OF BORON-TRIFLUORIDE AND HELIUM DETECTORS IN NEUTRON AND GAMMA-FIELDS, IEEE transactions on nuclear science, 45(3), 1998, pp. 636-642
Citations number
9
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
45
Issue
3
Year of publication
1998
Pages
636 - 642
Database
ISI
SICI code
0018-9499(1998)45:3<636:AMFCDO>2.0.ZU;2-T
Abstract
A method developed to measure the degradation of neutron detectors in neutron and gamma fields has been applied to five models of boron trif luoride (BF3) detectors from major suppliers, and to a special helium (He-3) detector model. The detectors tested all have about the same no minal thermal neutron sensitivity and overall dimensions. The results showed widely different neutron and gamma durability for BF3 models, a n undesirable time-dependent gamma degradation followed by recovery fo r some BF3 models, and very robust performance of the modified He-3 de tector.