DEVELOPMENT OF A PHOSWICH DETECTOR TO DETECT NEUTRONS AND CHARGED-PARTICLES FOR SPACE APPLICATION

Citation
M. Takada et al., DEVELOPMENT OF A PHOSWICH DETECTOR TO DETECT NEUTRONS AND CHARGED-PARTICLES FOR SPACE APPLICATION, IEEE transactions on nuclear science, 45(3), 1998, pp. 888-893
Citations number
8
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
45
Issue
3
Year of publication
1998
Pages
888 - 893
Database
ISI
SICI code
0018-9499(1998)45:3<888:DOAPDT>2.0.ZU;2-Q
Abstract
We have developed a new phoswich detector consisting of an NE213 liqui d scintillator surrounded by an NE115 plastic scintillator to measure high energy neutron spectra accurately in a charged-particle mixed fie ld. Differences in the decay time constants of light outputs from the two scintillators make it possible to separate charged-particle-induce d and neutron-induced pulses. Experiments were carried out at the RIKE N and at the NIRS cyclotrons. This detector was found to be able to di scriminate neutron and gamma-ray events from external-proton events. T his detector will be useful to measure high energy neutron in space.