A method to determine depth of interaction (DOI) from a PET detector m
odule is described and evaluated. The basic element of the DOI detecto
r module is a two crystal detector unit. The hypothesis is that by con
trolling how light is shared between two crystals (A and B) DOI inform
ation can be extracted from the ratio of light collected using simple
Anger type logic [A/(A+B)]. The interface between crystals is designed
so that a significant amount of light is shared when a photon interac
ts near the front face of a crystal and very little light is shared wh
en an interaction occurs near the back of a crystal. The effects of su
rface treatment (e.g., polished, roughened) and optical coupling compo
unds are investigated. EGO, GSO and LSO detector units have been evalu
ated. A DOI uncertainty; of similar to 6mm was attained for the front
section (similar to 4mm) of a 2x2x20mm LSO detector unit. A method to
decode a 64 crystal detector module (32 detector units) using a 16 cha
nnel multi-anode photomultiplier tube is described.