K. Ute et al., DIRECT DETERMINATION OF MOLECULAR-WEIGHT DISTRIBUTION BY SIZE-EXCLUSION CHROMATOGRAPHY WITH 750 MHZ H-1-NMR DETECTION (ONLINE SEC-NMR), Polymer Journal, 30(5), 1998, pp. 439-443
A size exclusion chromatograph coupled to a 750 MHz H-1 NMR spectromet
er (on-line SEC-NMR) was used to analyze molecular weight distribution
s of isotactic PMMAs with (M) over bar(n) ranging from 3.27 x 10(3) to
2.22 x 10(4). The molecular weight of the polymer in flowing eluate w
as determined directly (without a conventional calibration procedure)
from the relative intensity of NMR signals due to the end-group and re
peating units. The molecular weight distribution determined in this ma
nner was proved to be accurate by SEC-NMR experiments on uniform isota
ctic PMMAs consisting of exactly 23, 40, and 80 repeating units. Inves
tigation was also undertaken to optimize flow rate and NMR parameters
for quantitative SEC-NMR measurements. The signal-to-noise ratio of 75
0 MHz SEC-NMR spectra was 9.2 times that obtained previously with 500
MHz SEC-NMR operated under similar conditions.