A SIMPLE DESIGN METHODOLOGY FOR ELLIPTIC CROSS-SECTION, TRANSVERSE, ASYMMETRIC, HEAD GRADIENT COILS FOR MRI

Citation
S. Crozier et Dm. Doddrell, A SIMPLE DESIGN METHODOLOGY FOR ELLIPTIC CROSS-SECTION, TRANSVERSE, ASYMMETRIC, HEAD GRADIENT COILS FOR MRI, IEEE transactions on biomedical engineering, 45(7), 1998, pp. 945-948
Citations number
18
Categorie Soggetti
Engineering, Biomedical
ISSN journal
00189294
Volume
45
Issue
7
Year of publication
1998
Pages
945 - 948
Database
ISI
SICI code
0018-9294(1998)45:7<945:ASDMFE>2.0.ZU;2-9
Abstract
A simple design process for the design of elliptical cross-section, tr ansverse gradient coils for use in magnetic resonance imaging (MRI) is presented. This process is based on a flexible stochastic optimizatio n method and results in designs of high linearity and efficiency with low switching times. A design study of a shielded, transverse asymmetr ic elliptical coil set for use in neural imaging is presented and incl udes the minimization of the torques experienced by the gradient set.