Y. Chen et al., STUDIES ON MONOXIDE EMISSION-SPECTROMETRY OF RARE-EARTH ELEMENTS - PART-V - DETERMINATION OF DY IN RARE-EARTH CONCENTRATES BY DUAL-WAVELENGTH METHOD, Analyst, 123(6), 1998, pp. 1235-1237
The enhancement effect of Pr, Y, Eu and La on the determination of Dy
by air-acetylene flame monoxide flame emission spectrometry is reporte
d. The dual wavelength method (526.3 nm as the analytical wavelength a
nd 526.7 nm as the reference wavelength) was used to eliminate spectra
l interferences. A new method was established for direct determination
of Dy (about 4%) in rare-earth mixture concentrates using Pr as enhan
cing agent and chemical interference inhibitor. The detection limit is
0.4 mu g ml(-1) Dy and the linear range is; 5-900 mu g ml(-1) Dy.