EXPERIMENTAL-STUDY ON TEMPERATURE DISTRIBUTION WITHIN A WIDE-GAP CONTINUOUS FREE-FLOW ELECTROPHORESIS CHAMBER

Citation
Hb. Wang et al., EXPERIMENTAL-STUDY ON TEMPERATURE DISTRIBUTION WITHIN A WIDE-GAP CONTINUOUS FREE-FLOW ELECTROPHORESIS CHAMBER, Electrophoresis, 19(7), 1998, pp. 1231-1233
Citations number
6
Categorie Soggetti
Biochemical Research Methods","Chemistry Analytical
Journal title
ISSN journal
01730835
Volume
19
Issue
7
Year of publication
1998
Pages
1231 - 1233
Database
ISI
SICI code
0173-0835(1998)19:7<1231:EOTDWA>2.0.ZU;2-2
Abstract
The temperature distribution within a wide-gap (3 mm) continuous free- flow electrophoresis (CFFE) chamber was investigated by direct measure ments. The walls of the chambers consisted of Plexiglas and thermocoup les were used as temperature sensors. The study of different operating and heat dissipating conditions revealed that lateral temperature gra dients arise in CFFE chambers in addition to an axial increase in temp erature of 0.25 degrees C per Watt of electric power and per minute of residence time in the absence of cooling.