Re. Davidsen et Sw. Smith, A 2-DIMENSIONAL ARRAY FOR B-MODE AND VOLUMETRIC IMAGING WITH MULTIPLEXED ELECTROSTRICTIVE ELEMENTS, Ultrasonic imaging (Print), 19(4), 1997, pp. 235-250
Citations number
24
Categorie Soggetti
Acoustics,"Radiology,Nuclear Medicine & Medical Imaging","Engineering, Biomedical
A 2:1 multiplexed 2-D array has been developed that has a sparse eleme
nt pattern designed for real time volumetric imaging and an alternate
element pattern designed for B-mode imaging. For volumetric imaging, a
small aperture was used to provide a wide transmit beam, allowing mul
tiple beams to be received simultaneously. For B-mode imaging, a large
r aperture with a more narrow transmit beam was used to improve image
quality. Sparse random element patterns were evaluated by beamplot and
cyst image simulations. Using the alternate element pattern for B-mod
e imaging, simulated cyst contrast was improved by 28 %. The multiplex
ed transducer was fabricated using an electrostrictive material in whi
ch array elements were activated and deactivated by a de bias field. T
he transducer had a 3.4 MHz center frequency with 46 % bandwidth, whic
h was consistent with KLM simulations. The high dielectric constant of
the electrostrictive material resulted in an element clamped capacita
nce of 14.3 pF versus 2 pF for a convention PZT element. An output off
isolation of -35 dB was measured for transmit and -61 dB for receive.
The array was integrated with the volumetric scanner and used to make
real time images oi a cyst phantom. The images showed improved cyst c
ontrast using the alternate aperture for B-mode imaging. Copyright 199
8 by Dynamedia, Inc.