COMPARISON OF A MULTIPORE COLUMN WITH A MIXED-BED COLUMN FOR SIZE-EXCLUSION CHROMATOGRAPHY

Citation
M. Nagata et al., COMPARISON OF A MULTIPORE COLUMN WITH A MIXED-BED COLUMN FOR SIZE-EXCLUSION CHROMATOGRAPHY, Journal of liquid chromatography & related technologies, 21(10), 1998, pp. 1471-1484
Citations number
6
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
ISSN journal
10826076
Volume
21
Issue
10
Year of publication
1998
Pages
1471 - 1484
Database
ISI
SICI code
1082-6076(1998)21:10<1471:COAMCW>2.0.ZU;2-7
Abstract
A new type of column packed with polystyrene gels which have a broad p ore size distribution in a particle thenceforth referred to as the mul tipore column) was developed, and its performance was compared with th e so-called linear column thenceforth referred to as the mixed-bed col umn). The chromatogram of the epoxy resin (Epikote 1009) on the multip ore column showed a monomodal distribution, but the chromatogram on th e mixed-bed column showed the inflection on the chromatogram at the mo lecular mass of about 10,000 as polystyrene equivalent. There was a bi g difference in the pore size distributions of gels packed in the two types of columns; the pore size distribution of the multipore column s howed the trapezoid, while that of the mixed-bed column had sharp maxi ma for pores with a diameter of 0.08 mu m. This difference in the pore size distribution may be the main reason for the inflection phenomeno n, though the: calibration curves of these two types of columns were a lmost the same, The multipore column also had the advantage of the mix ed-bed column in the linearity of the calibration curve.