M. Nagata et al., COMPARISON OF A MULTIPORE COLUMN WITH A MIXED-BED COLUMN FOR SIZE-EXCLUSION CHROMATOGRAPHY, Journal of liquid chromatography & related technologies, 21(10), 1998, pp. 1471-1484
Citations number
6
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A new type of column packed with polystyrene gels which have a broad p
ore size distribution in a particle thenceforth referred to as the mul
tipore column) was developed, and its performance was compared with th
e so-called linear column thenceforth referred to as the mixed-bed col
umn). The chromatogram of the epoxy resin (Epikote 1009) on the multip
ore column showed a monomodal distribution, but the chromatogram on th
e mixed-bed column showed the inflection on the chromatogram at the mo
lecular mass of about 10,000 as polystyrene equivalent. There was a bi
g difference in the pore size distributions of gels packed in the two
types of columns; the pore size distribution of the multipore column s
howed the trapezoid, while that of the mixed-bed column had sharp maxi
ma for pores with a diameter of 0.08 mu m. This difference in the pore
size distribution may be the main reason for the inflection phenomeno
n, though the: calibration curves of these two types of columns were a
lmost the same, The multipore column also had the advantage of the mix
ed-bed column in the linearity of the calibration curve.