Md. Bi et al., A VALID HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHOD FOR QUANTIFICATION OF LUTEINIZING-HORMONE-RELEASING HORMONE, Journal of liquid chromatography & related technologies, 21(10), 1998, pp. 1503-1509
Citations number
5
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
An isocratic technique was developed for the analysis of luteinizing h
ormone releasing hormone (LHRH) by high performance liquid chromatogra
phy (HPLC) using 210 nm UV detection, ZORBAX ODS column (4.6 mm x 15 c
m), mobile phase (86% triethylamine phosphate buffer: 14% acetonitrile
), and 1.5 mL/min flow rate. The coefficient of variation (C.V.) f'or
precision and proportionality assays was lower than 5% for all concent
rations studied. The detection limit of LHRH was 1 ng/mL.