GAS-CHROMATOGRAPHIC DETERMINATION OF ZOPICLONE IN PLASMA AFTER SOLID-PHASE EXTRACTION

Citation
Y. Gaillard et al., GAS-CHROMATOGRAPHIC DETERMINATION OF ZOPICLONE IN PLASMA AFTER SOLID-PHASE EXTRACTION, Journal of chromatography. Biomedical applications, 619(2), 1993, pp. 310-314
Citations number
12
Categorie Soggetti
Chemistry Analytical
ISSN journal
03784347
Volume
619
Issue
2
Year of publication
1993
Pages
310 - 314
Database
ISI
SICI code
0378-4347(1993)619:2<310:GDOZIP>2.0.ZU;2-M
Abstract
A gas chromatographic technique for determining zopiclone based on a s olid-phase extraction procedure with C18 cartridge for sample clean-up is presented. Quantification can be achieved with 1 ml of plasma. The method uses prazepam as internal standard. Zopiclone is separated on a 5% phenyl methyl silicone analytical column and detected with an ele ctron-capture detector, which consequently allows a limit of quantitat ion of 2 mug/l. It is thus simple, rapid, sensitive and linear over th e range 5-2000 mug/l.