TRANSCATHETER OCCLUSION OF PATENT DUCTUS-ARTERIOSUS WITH A NEW DETACHABLE COIL SYSTEM (DUCTOCCLUD) - A MULTICENTER CLINICAL-TRIAL

Citation
S. Oho et al., TRANSCATHETER OCCLUSION OF PATENT DUCTUS-ARTERIOSUS WITH A NEW DETACHABLE COIL SYSTEM (DUCTOCCLUD) - A MULTICENTER CLINICAL-TRIAL, Japanese Circulation Journal, 62(7), 1998, pp. 489-493
Citations number
25
Categorie Soggetti
Peripheal Vascular Diseas
ISSN journal
00471828
Volume
62
Issue
7
Year of publication
1998
Pages
489 - 493
Database
ISI
SICI code
0047-1828(1998)62:7<489:TOOPDW>2.0.ZU;2-P
Abstract
A multicenter clinical trial of DuctOcclud, a new detachable coil for transcatheter occlusion of patent ductus arteriosus (PDA), was conduct ed. DuctOcclud was used in 35 patients (12 male and 23 female) for tra nscatheter occlusion of PDA between January, 1996, and April, 1997, Th e age of the patients ranged from 0.5 to 27.2 years (median 7.6 years) and weight from 6.3 to 70.0kg (median 23,0kg). The smallest diameter of PDA was 2.0+/-0.7 mm (range 1.0-3.3 mm). Pulmonary-systemic flow ra tio (Qp/Qs) was 1.3+/-0.3 (range 1.0-2.2), The coils were successfully implanted in 32 (91%) patients. Of 31 patients who were followed 6 mo nths after the procedure, 26 (84%) had no residual shunt and 5 (16%) h ad trivial residual shunt. One patient had infective endocarditis 1 mo nth after the procedure but recovered completely, There were no incide nces of coil embolization, hemolysis, late coil migration, or pulmonar y artery stenosis. We conclude that DuctOcclud is a safe and effective device of transcatheter occlusion of PDA.