Wj. Hurst et Ra. Martin, HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF OCHRATOXIN-AIN ARTIFICIALLY CONTAMINATED COCOA BEANS USING AUTOMATED SAMPLE CLEANUP, Journal of chromatography, 810(1-2), 1998, pp. 89-94
Citations number
18
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A HPLC method is described for the analysis of ochratoxin A at low-ppb
levels in samples of artificially contaminated cocoa beans. The sampl
es are extracted in a mixture of methanol-water containing ascorbic ac
id, adjusted to pH and evaporated to dryness. Samples in this state ar
e then placed onto a Benchmate sample preparation workstation where C-
18 solid-phase extraction operations are performed. The resulting mate
rials are evaporated to dryness and analyzed by reversed-phase HPLC wi
th fluorescence detection. The method was evaluated for accuracy and p
recision with R.S.D.s for multiple injections of sample and standard c
alculated to be 1.1% and 2.5% for sample and standard, respectively. R
ecoveries of ochratoxin A added to cocoa beans ranged from 87-106% ove
r the range of the assay. (C) 1998 Elsevier Science B.V. All rights re
served.