X-RAY-ANALYSIS OF LIGHT-ELEMENTS BY MONOC HROMATIC LIGHT

Citation
Sj. Liu et al., X-RAY-ANALYSIS OF LIGHT-ELEMENTS BY MONOC HROMATIC LIGHT, Gaoneng wuli yu he wuli, 22(4), 1998, pp. 378-382
Citations number
10
Categorie Soggetti
Physics, Particles & Fields","Physics, Nuclear
Journal title
ISSN journal
02543052
Volume
22
Issue
4
Year of publication
1998
Pages
378 - 382
Database
ISI
SICI code
0254-3052(1998)22:4<378:XOLBMH>2.0.ZU;2-C
Abstract
As variable energy monochromatic sources both synchrotron radiation mo nochromatic light and proton induced X-ray emission were used to study X-ray fluorecence of K, Ca, Ti light elements. The results indicate t hat this method not only reduces the bremsstrahlung background but als o eliminates the interference with some other elements by selecting ap propriate exciting energy. Thus the analytic sensitivity of K, Ca, Ti elements are considerably improved.