OPTIMIZATION OF PT SENSITIVITY IN PIXE ANALYSIS OF THIN BIOLOGICAL SAMPLES

Citation
Tb. Tiourina et al., OPTIMIZATION OF PT SENSITIVITY IN PIXE ANALYSIS OF THIN BIOLOGICAL SAMPLES, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 139(1-4), 1998, pp. 186-191
Citations number
8
Categorie Soggetti
Instument & Instrumentation","Nuclear Sciences & Tecnology","Physics, Atomic, Molecular & Chemical","Physics, Nuclear
ISSN journal
0168583X
Volume
139
Issue
1-4
Year of publication
1998
Pages
186 - 191
Database
ISI
SICI code
0168-583X(1998)139:1-4<186:OOPSIP>2.0.ZU;2-U
Abstract
To obtain Pt concentration distributions in tumor biopsies, micro-PIXE is used. The Limit of Detection (LOD) for Pt is determined by the bac kground production cross section together with the production cross se ction for the Pt-L lines. Both are functions of the type and energy of the projectile. In order to optimise the experimental conditions, the proton energy is varied from 2.5 to 5 MeV and the corresponding X-ray spectra are analysed. The optimum LOD is about 1 mu g/g for a sample thickness of 1 mg/cm(2) and a total charge of 100 mu C. A study is per formed to investigate the possibility of elemental analysis by means o f alpha excitation. The corresponding LOD is 6 mu g/g with 200 mu C ch arge (He++ was used). For alpha excitation, sample damage limits the b eam current and, thus, causes a tremendous increase in the measurement time. (C) 1998 Published by Elsevier Science B.V.