Hj. Mascher et C. Kikuta, DETERMINATION OF NORFLOXACIN IN HUMAN PLASMA AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND FLUORESCENCE DETECTION, Journal of chromatography, 812(1-2), 1998, pp. 381-385
Citations number
9
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A method for the determination of norfloxacin in human plasma and urin
e is described. Plasma samples were deproteinized using acetonitrile.
The supernatant was analysed by C-18 HPLC. Fluorescence detection at a
n excitation wavelength of 300 nm and an emission wavelength of 450 nm
was utilized. The assay was validated in the concentration range of 3
1 to 2507 ng/ml when 0.5-ml aliquots of plasma were handled. The intra
-day precision of the spiked quality control samples ranged from +/-0.
37 to +/-4.14% in plasma (concentration range: 70.3-2109.2 ng/ml) and
from +/-0.51 to +/-1.56% in urine (concentration range: 7.5-299.4 mu g
/ml). The intra-day accuracy obtained for norfloxacin in the quality c
ontrol samples ranged from -5.18% to -9.47% in plasma and from -10.56%
to -5.91% in urine. The assay has been used to support human pharmaco
kinetic studies. (C) 1998 Published by Elsevier Science B.V. All right
s reserved.