IN-SITU RHEOMETRIC SHEARING APPARATUS AT THE NIST CENTER FOR NEUTRON RESEARCH

Citation
Gc. Straty et al., IN-SITU RHEOMETRIC SHEARING APPARATUS AT THE NIST CENTER FOR NEUTRON RESEARCH, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 408(2-3), 1998, pp. 511-517
Citations number
31
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
408
Issue
2-3
Year of publication
1998
Pages
511 - 517
Database
ISI
SICI code
0168-9002(1998)408:2-3<511:IRSAAT>2.0.ZU;2-Y
Abstract
A constant stress Couette cell rheometer has been modified to measure simultaneously the viscometric properties and neutron scattering inten sities from the NIST Center for Neutron Research (NCNR) small angle ne utron scattering (SANS) spectrometers. The apparatus was constructed t o complement scattering radiation experiments of solutions under shear . The apparatus is capable of high accuracy measurements with viscosit ies that can range over about ten orders of magnitude. Sample plots sh owing simultaneous measurements of the viscosity and the anisotropic s cattering patterns from micellar solutions are presented. The apparatu s is easily adapted for light scattering. (C) 1998 Published by Elsevi er Science B.V. All rights reserved.