ENLARGED LINEARITY RANGE OF A RASNIK SYSTEM USING AN INTEGRATING SPHERE

Citation
J. Alberdi et al., ENLARGED LINEARITY RANGE OF A RASNIK SYSTEM USING AN INTEGRATING SPHERE, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 408(2-3), 1998, pp. 591-593
Citations number
5
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
408
Issue
2-3
Year of publication
1998
Pages
591 - 593
Database
ISI
SICI code
0168-9002(1998)408:2-3<591:ELROAR>2.0.ZU;2-Y
Abstract
Rasniks elements are foreseen to be used in the CMS alignment system. The large displacements and deformations expected for the different su bdetectors demands large linearity response range of the alignment com ponents. By the use of an integrating sphere coupled to the light sour ce, we have optimized the homogenity of the beam, leading to improveme nt, by factor three, in the linearity range of the Rasnik system with respect to conventional configurations, preserving the measurement pos ition resolution. The range obtained coincides with the maximum reacha ble with the geometry of the used components. (C) 1998 Elsevier Scienc e B.V. All rights reserved.