DETERMINATION OF IRON-ION UTILIZING POLY (P-VINYLBENZOYLACETONE) MODIFIED MICROELECTRODE

Citation
J. Motonaka et al., DETERMINATION OF IRON-ION UTILIZING POLY (P-VINYLBENZOYLACETONE) MODIFIED MICROELECTRODE, Analytica chimica acta, 369(1-2), 1998, pp. 87-92
Citations number
14
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
369
Issue
1-2
Year of publication
1998
Pages
87 - 92
Database
ISI
SICI code
0003-2670(1998)369:1-2<87:DOIUP(>2.0.ZU;2-W
Abstract
A porous micro-electrode with immobilized poly (p-vinylbenzoylacetone) (PVBA) was developed for measuring iron ion. A porous carbon material was packed into the tip of the platinum micro-electrode with a cavity at the top. The porous microelectrode was dipped in the coating tetra hydrofuran solution with PVBA and di-2-ethylhexyl phthalate and dried overnight at room temperature. Calibration curve for the measurement o f iron, the effects of preconcentration condition, scan rate, pH, and temperature, and lifetime of the PVBA modified micro-electrode were ex amined. Under optimal conditions, the response of the sensor was linea r in the concentration range of 0.01-1 mM for iron ion. (C) 1998 Elsev ier Science B.V. All rights reserved.