CLINICALLY RELEVANT DISCOVERY RECOGNIZED DURING SIMULATOR TRAINING SESSION - FREESTANDING PEEP VALVE SOLVES PROBLEM OF INCOMPETENT EXHALATION UNIDIRECTIONAL VALVE IN AN ANESTHESIA CIRCLE BREATHING SYSTEM

Citation
Ty. Euliano et al., CLINICALLY RELEVANT DISCOVERY RECOGNIZED DURING SIMULATOR TRAINING SESSION - FREESTANDING PEEP VALVE SOLVES PROBLEM OF INCOMPETENT EXHALATION UNIDIRECTIONAL VALVE IN AN ANESTHESIA CIRCLE BREATHING SYSTEM, JOURNAL OF CLINICAL MONITORING AND COMPUTING, 14(3), 1998, pp. 191-194
Citations number
6
Categorie Soggetti
Anesthesiology,"Medical Informatics
ISSN journal
13871307
Volume
14
Issue
3
Year of publication
1998
Pages
191 - 194
Database
ISI
SICI code
1387-1307(1998)14:3<191:CRDRDS>2.0.ZU;2-#
Abstract
While conducting a case-based teaching session using a full-scale simu lator, the authors discovered that placement of a unidirectional PEEP valve in the expiratory limb corrects the gas flow pattern, despite an exhalation valve failure.