I. Kheterpal et al., A 3-WAVELENGTH LABELING APPROACH FOR DNA-SEQUENCING USING ENERGY-TRANSFER PRIMERS AND CAPILLARY-ELECTROPHORESIS, Electrophoresis, 19(8-9), 1998, pp. 1403-1414
Citations number
41
Categorie Soggetti
Biochemical Research Methods","Chemistry Analytical
Capillary electrophoresis DNA sequencing has been accomplished by usin
g four different energy transfer primers and three fluorescence detect
ion channels. Methods have also been developed to deconvolve the three
-color data into the four base concentrations. The nonnegative least s
quares and model selection method resulted in the best accuracy. The t
hree-color data were compared to sequencing data obtained using four d
etection channels and four energy transfer primers. The average accura
cy rates obtained over three 500 base M13mp18 runs using three-color c
oding were 96% including 18 uncallable compressions and 99.6% if these
compressions are excluded. The average accuracy rate obtained using f
our-color coding was 96.3% including 18 uncallable compressions and 99
.9% if these compressions are excluded. Although it is unlikely that t
hree-color schemes will replace four-color sequencing, these methods h
ave exposed basic concepts that will be useful in the development of h
igher-order multiplex coding methods for DNA analysis.