Mtsr. Gomes et al., DETERMINATION OF CYANIDE IN WASTE-WATERS USING A QUARTZ-CRYSTAL MICROBALANCE, Sensors and actuators. B, Chemical, 48(1-3), 1998, pp. 383-386
A method based on a quartz crystal microbalance (QCM) was used for the
first time to analyse cyanide in real samples. The method was applied
to the analysis of waste waters from an electroplating industry. The
analysis of replicates of five different samples ranging from 0.931 to
853 mg l(-1) showed relative standard deviations between 1.7% and 9.3
%. There was no evidence of systematic differences between the two set
s of results obtained by the QCM method or by the ASTM method. However
, the QCM method was faster and needed a sample volume 100 times small
er than the ASTM method. (C) 1998 Elsevier Science S.A. All rights res
erved.