DETERMINATION OF TRACE EUROPIUM BASED ON NEW FLUOROMETRIC SYSTEM OF EUROPIUM(III) WITH THENOYLTRIFLUOROACETONE AND N'-DINAPHTHYL-N,N'-DIPHENYL-3,6-DIOXAOCTANEDIAMIDE
W. Yang et al., DETERMINATION OF TRACE EUROPIUM BASED ON NEW FLUOROMETRIC SYSTEM OF EUROPIUM(III) WITH THENOYLTRIFLUOROACETONE AND N'-DINAPHTHYL-N,N'-DIPHENYL-3,6-DIOXAOCTANEDIAMIDE, Talanta, 46(4), 1998, pp. 527-532
In the present paper, N'-dinaphthyl-N,N'-diphenyl-3,6-dioxaoctanediami
de acts as a specific reagent for enhancing the fluorescence intensity
of Eu(III) complex with thenoyl trifluoroacetone (TTA), the spectrofl
uorimetric determination of trace amounts of Eu(III) based on the abov
e system was carried out and its luminescence mechanism was studied. T
he excitation and emission wavelengths are 343.6 and 613.3 nm, respect
ively. The fluorescence intensities vary linearly with the concentrati
on of europium(III) in the range 3.647 x 10(-3)-3.039 mu g ml(-1) for
the original fluorescence with a detection limit down to 2.279 x 10(-4
) mu g ml(-1) and the standard deviation is 0.063 mu g ml(-1) for 10 t
imes measurements? and in the range 7.598 x 10(-4)-0.0243 mu g ml(-1)
(SD = 0.035 for 15 times measurements), 0.06078-0.6100 mu g ml(-1) (SD
= 0.52 for 10 times measurements) for the first derivative fluorescen
ce signal with a detection limit down to 8.566 x 10(-5) mu g ml(-1). T
he interferences of other rare earths and some of inorganic ions are d
escribed. This method is a direct, rapid, selective and sensitive anal
ytical method for the determination of trivalent europium in rare eart
h ore samples and high purity of rare earth oxides. (C) 1998 Elsevier
Science B.V. All rights reserved.