Cj. Bruegge et al., MISR PRELAUNCH INSTRUMENT CALIBRATION AND CHARACTERIZATION RESULTS, IEEE transactions on geoscience and remote sensing, 36(4), 1998, pp. 1186-1198
Each of the nine cameras that compose the Multiangle Imaging SpectroRa
diometer (MISR) has been rigorously tested, characterized, and calibra
ted. Requirements on these tests include a 3% (1 sigma) radiometric ca
libration requirement, spectral response function determination of bot
h the in- and out-of-band regions, and distortion mapping. The latter
test determines the relative look-angle to the ground corresponding to
each focal plane detector element. This is established to within one-
tenth of the instantaneous field-of-view. Most of the performance test
ing was done on the cameras as they completed assembly. This was done
to take advantage of the serial delivery of the hardware, minimize the
required size of the thermal-vacuum facilities, and allow testing to
occur early in the schedule allocated for the hardware build. This pro
ved to be an effective strategy, as each of the test objectives was me
t, Additional testing as an integrated instrument included verificatio
n of the data packetization, camera pointing, and clearances of the fi
elds-of-view. Results of these studies have shown that the MISR camera
s are of high quality and will meet the needs of the MISR science comm
unity. Highly accurate calibration data are on-hand and available for
conversion of camera output to radiances.