B. Faddegon et al., CLINICAL CONSIDERATIONS OF MONTE-CARLO FOR ELECTRON RADIOTHERAPY TREATMENT PLANNING, Radiation physics and chemistry (1993), 53(3), 1998, pp. 217-227
Citations number
36
Categorie Soggetti
Nuclear Sciences & Tecnology","Chemistry Physical","Physics, Atomic, Molecular & Chemical
Technical requirements for Monte Carlo based electron radiotherapy tre
atment planning are outlined. The targeted overall accuracy for estima
te of the delivered dose is the least restrictive of 5% in dose, 5 mm
in isodose position. A system based on EGS4 and capable of achieving t
his accuracy is described. Experience gained in system design and comm
issioning is summarized. The key obstacle to widespread clinical use o
f Monte Carlo is lack of clinically acceptable measurement based metho
dology for accurate commissioning. (C) 1998 Elsevier Science Ltd. All
rights reserved.