Dm. Katz et al., QUANTIFICATION OF HOLOGRAPHIC FRINGE DATA - COMPARISON OF INTACT AND IMPLANTED FEMURS, Medical engineering & physics, 20(2), 1998, pp. 114-123
The strains developed on the surface of a femur with and without an im
planted femoral hip replacement were investigated using holographic in
terferometry (HI). Interferometric fringe data from the four aspects o
f the loaded femur were quantified and the resulting curvature calcula
ted. Curvature was related to the strain along the long axis of the fe
mur and the results for the intact and implanted femur compared. Chang
es were found in femoral strain patterns resulting from the implantati
on of both a cemented and an uncemented Norwich femoral component. All
four aspects of the proximal 120 mm of the femur showed reduced strai
n as a result of prosthesis implantation irrespective of the method of
prosthesis fixation. The distal half of the femur showed no such chan
ges, No statistical difference was found between the loading patterns
in the proximal femur of the uncemented and the cemented prosthesis. (
C) 1998 IPEM. Published by Elsevier Science Ltd. All rights reserved.