AN EVALUATION OF THE CAMINO VENTRICULAR BOLT SYSTEM IN CLINICAL-PRACTICE

Citation
Ir. Chambers et al., AN EVALUATION OF THE CAMINO VENTRICULAR BOLT SYSTEM IN CLINICAL-PRACTICE, Neurosurgery, 33(5), 1993, pp. 866-868
Citations number
5
Categorie Soggetti
Surgery,Neurosciences
Journal title
ISSN journal
0148396X
Volume
33
Issue
5
Year of publication
1993
Pages
866 - 868
Database
ISI
SICI code
0148-396X(1993)33:5<866:AEOTCV>2.0.ZU;2-9
Abstract
THE CAMINO VENTRICULAR bolt system has been used to monitor intracrani al pressure in patients after severe head injury. The correlation betw een the ventricular pressure measured with the Camino device and an ex ternal transducer showed that the Camino accurately measured intracran ial pressure over a wide range, but that it read an average of 1.15 mm Hg higher than that obtained by the external transducer. The techniqu e has the advantage over a remote transducer because it is sited withi n the ventricle. This may be of value in wave-form analysis.