ACTIVITY MEASUREMENT OF AU-198 IN GOLD FOILS

Citation
Te. Sazonova et al., ACTIVITY MEASUREMENT OF AU-198 IN GOLD FOILS, Applied radiation and isotopes, 49(9-11), 1998, pp. 1185-1189
Citations number
2
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging","Chemistry Inorganic & Nuclear
Journal title
Applied radiation and isotopes
ISSN journal
09698043 → ACNP
Volume
49
Issue
9-11
Year of publication
1998
Pages
1185 - 1189
Database
ISI
SICI code
0969-8043(1998)49:9-11<1185:AMOAIG>2.0.ZU;2-D
Abstract
A new method is reported for the activity measurement of (198)AU in go ld foils. The coincidence counting equations are examined. From these the experimental conditions necessary to ensure accurate results are o btained. These conditions are achieved by careful choice of NaI(TI) cr ystals and gamma-ray absorbers. The uncertainty components amounted to 0.15% for type A and 0.3% for type B. (C) 1998 Published by Elsevier Science Ltd. All rights reserved.